2.4 Making Connections
38
<Connecting to the terminals of the Universal Unit>
Required items: phillips screwdriver (for M3 screws), flat-blade screwdriver (tip width 2.6
mm), RTD
Recommended wire diameter:Single strand diameter:
φ
0.4 mm -
φ
1.2 mm (AWG26-16)
Multi-strand:
0.2 mm
2
- 0.75 mm
2
(AWG24-20)
Standard insulation stripping length: 8 mm
Temperature Measurement with a Resistance Temperature
Detector (RTD)
1
With a Phillips screwdriver,
loosen the screws in the termi-
nal block cover, and lift the
cover.
2
Push in the button on the con-
nector with a flatblade screw-
driver.
3
With the button held in, insert
the cable into the cable con-
nection hole.
6
Replace the terminal block
cover, and tighten the screws.
Torque screws to no more than
0.6 N
•
m.
4
Release the button.
The cable is locked.
4-wire configuration
3-wire configuration
2
3
4
button
5
5
Connect to the measurement
object.
The insulation colors of wires, which are to be
connected to the plus and minus terminals, differ
depending on countries or manufactures. Check
with the manufacturer on the colors and connect
them according to instructions.
Summary of Contents for LR8400-20
Page 1: ......
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 14...
Page 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Page 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Page 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Page 238: ...10 8 About Communications Commands 226...
Page 256: ...11 3 LR8501 Universal Unit Specifications 244...
Page 262: ...12 3 Disposing of the HiLogger 250...
Page 296: ...Index Index 4...
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......