2.4 Making Connections
36
< Voltage/Temp Unit >
Required items: phillips screwdriver (for M3 screws), measurement leads (for voltage
measurement), or thermocouple (for temperature measurement)
If connecting crimped terminals to the analog input terminals, use insulated terminals
for M3 screws with the size shown below.
6 mm
max.
6 mm
max.
Voltage and Thermocouple Temperature Measurement
1
Loosen the screws in the ter-
minal block cover, and lift the
cover.
2
Loosen the terminal block
screws, insert the bare ends of
the leads, and tighten the
screws.
3
Connect to the measurement
object.
4
Replace the terminal block
cover, and tighten the screws.
Torque screws to no more than
0.6 N
•
m.
When measuring instrumentation
control equipment
(Applying 4 - 20 mA input current)
For details on in-
strumentation con-
trol equipment
measurements,
see (p. 57).
Connect a 250
Ω
shunt resistor as
shown.
2
3
The insulation colors of wires, which are to be
connected to the plus and minus terminals, differ
depending on countries or manufactures. Check
with the manufacturer on the colors and connect
them according to instructions.
K thermocouples experience a physicalphenomenon known as short
range ordering, which can prevent accurate measurement within the
range of 250°C to 600°C.
Please choose a sensor after checking with the manufacturer of the
thermocouple you intend to use.
Summary of Contents for LR8400-20
Page 1: ......
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 14...
Page 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Page 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Page 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Page 238: ...10 8 About Communications Commands 226...
Page 256: ...11 3 LR8501 Universal Unit Specifications 244...
Page 262: ...12 3 Disposing of the HiLogger 250...
Page 296: ...Index Index 4...
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......