Appendix 10 Introduction to Measurement Applications
A
27
Appendix
Setting Example
Unit: PLS&ALM, Channel: P1,
Input: Count, Count Mo: Add
Filter: On
Scaling
Cond: Dec, Unit: kW, 1kW=50,000 Pulse
Disp Span
Type: Up/Lwr, Upper: 10000, Lower: 0
Make input channel settings on the
[CH]
screen.
Measurement Hints
• Enable scaling to read measured values directly as electrical energy [kWh]. The display range
changes automatically when scaling is enabled, so set the upper and lower display limits after
enabling scaling.
• When the watt-hour meter uses mechanical (relay) contacts for output, chattering effects can
be removed by enabling the filter.
0 kW
10000 kW
When set as shown at the left,
the vertical axes appear like this.
1
2
Select
Apply
3
Start and finish measuring
• Data is recorded for the specified recording time and saved
to CF card. Recording stops 30 days after starting.
• To stop recording early, press
STOP
.
Press
START
.
Summary of Contents for LR8400-20
Page 1: ......
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 14...
Page 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Page 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Page 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Page 238: ...10 8 About Communications Commands 226...
Page 256: ...11 3 LR8501 Universal Unit Specifications 244...
Page 262: ...12 3 Disposing of the HiLogger 250...
Page 296: ...Index Index 4...
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......