TOUCH SCREEN PAPERLESS RECORDER 

MODEL : ARC2020  

 

 

ACCSYS ELECTRONICS

 

140/6B, GOLDEN INDUSTRIAL ESTATE, JAWAHARLAL NEHRU ROAD,  

GERUGAMBAKKAM, CHENNAI - 600122 

Tel: 044 60505599 / 60505511 

E-mail : [email protected] 

 

Summary of Contents for ARC2020

Page 1: ...nics com Page 1 USER MANUAL MULTI COLOR TOUCH SCREEN PAPERLESS RECORDER MODEL ARC2020 ACCSYS ELECTRONICS 140 6B GOLDEN INDUSTRIAL ESTATE JAWAHARLAL NEHRU ROAD GERUGAMBAKKAM CHENNAI 600122 Tel 044 6050...

Page 2: ...ng INSTALLATION Installation Location Install the control panel in a location that meets the following criteria 1 Little or no mechanical vibration 2 No corrosive gases 3 Minimal temperature fluctuati...

Page 3: ...o 50 ohm Open and short circuit detection Type Range C Accuracy C 100 ohm Pt 385 200 to 400 C 0 2 C Recording Recording Rates User programmable from 1 sample per second to 1 sample every 10 hours Data...

Page 4: ...her Capacity Memory Internal memory 2 4 GB for Higher Data Storage Recording data on storage media Using the Setup menu the user can log the data in internal memory If this memory is full the logged d...

Page 5: ...11 Compliant Software Web server Based Software Safety and Environmental Operating Range 0 C to 60 C 10 to 90 RH non condensing Dimension Bezel dimension 195 H x 300 W x 5mm above panel Unit dimension...

Page 6: ...TOUCH SCREEN PAPERLESS RECORDER www accsyselectronics com Page 6...

Page 7: ...nics com Page 7 OPERATING PARAMETER SETTING Switch ON the device the following display will show the channel readings It will take a few seconds CONFIGURATION SETTING Click CONFIG in the display Enter...

Page 8: ...DER www accsyselectronics com Page 8 After entering the password the setting screen will be displayed as shown below TIME DATE SETTINGS In Setting SET Real Time Clock Set the real time date details Cl...

Page 9: ...w accsyselectronics com Page 9 LOG TIME SETTINGS In Setting SET Log Time Set the Log Time Of The Logger Click Ok to save the settings PASSWORD SETTINGS In Setting Click SET Password this feature is us...

Page 10: ...rm range from the specified values Low 999 High 9999 RANGE Set the range with respect to the Sensor type specification REF PAGE NO 3 OFFSET Set the offset range from 999 to 999 Default set 0 DECIMAL S...

Page 11: ...front panel of the Logger In Setting Click COPY Memory this feature is used to copy the Logger memory in to the Pen Drive If USB 2 0 drive is not inserted in the drive an error screen PEN DRIVE PROBL...

Page 12: ...an Ink jet printer through the PRINTER PORT provided in the back panel of Logger In Setting Click PRINT Memory the entire memory will be printed GRAPH After completing all the Settings Click HOME ico...

Page 13: ...ERMINAL DETAILS ID NUMBER In Setting ID NUMBER SETTING for Multi Logger Software DATA LOGGING OFF Logging Off The Data Logger Click HOME icon in the Setting screen In Setting DATA LOGGING OFF Input Te...

Page 14: ...n go For step 2 If no check for the power input connections check whether the unit is getting proper supply at proper terminals Connecting the supply to the wrong terminals may damage the unit permane...