77
Setting Threshold Values for the Comparator
Examples of how to read the comparator output table
Example 1:
If the measured resistance is less than or equal to the resistance
warning threshold, and the measured voltage is more than or equal
to the voltage warning threshold, the
[PASS]
segment will appear.
Example 2:
If the measured resistance exceeds the resistance warning
threshold but is less than or equal to the resistance rejection value,
and the measured voltage exceeds the voltage warning threshold,
the instrument will show the
[WARNING]
segment and produce
beeps.
When the resistance warning threshold and resistance are set to the same
value, the boundary conditions are as shown below:
Resistance
PASS
≤
Resistance
warning threshold =
Resistance
rejection value <
Resistance
FAIL