Error Messages
10.3 Error Messages
If the display shows an error, follow a procedure described below to
troubleshoot the error.
Message
Description
Solution
Error
Function error
(Zero adjustment
has ended in
failure)
Connect the test lead correctly
and perform zero adjustment.
Error no AdJ
Adjustment
data error (No-
adjustment error)
The instrument needs to
be repaired. Contact your
authorized Hioki distributor or
reseller.
Error Adc
A/D converter
communications
error
Error 001
Error 002
From
Error 011
to
Error 018
Internal variable
error
Error 008
Z3210
communications
error
(Poor connection,
malfunction of
the Z3210 or
hardware)
Replace the Z3210 with
another one. If the instrument
can communicate after the
replacement, the Z3210 is
faulty.
If the instrument cannot
communicate after the
replacement, the instrument is
faulty. Contact your authorized
Hioki distributor or reseller.