Measurement
To ensure safe operation, be sure to read “Operating Precautions”
(p. 11) before starting the measurements.
•
There is a considerable difference in the battery’s internal
resistances between a full-charge state and a discharge state.
To improve evaluation accuracy, perform measurements under
a constant condition (for example, in a full-charge state).
• Terminals of lead-acid batteries (objects under measurement)
have a high resistance. Thus, resistance values may differ
between contact locations, the case side and the tip of the
terminals. Place the test lead into contact with the terminals at
the same location every time.
See “11.4 Effects of Current Density” (p. 157).
•
Use the 9460 Clip Type Lead with Temperature Sensor
(option) to measure the temperature of the battery terminals.
Otherwise, use a non-contact thermometer, such as a radiation
thermometer, for safety.
•
If terminals are covered with an insulating layer, the
measurement current cannot flow sufficiently, resulting in
measurement failure. In such a case, clean the terminals to
remove the insulation layer before measurement.
3 Measurement
HIOKI BT3554F961-00
HIOKI BT3554F961-00
Summary of Contents for BT3554-50
Page 2: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Page 14: ...vi Contents HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Page 94: ...80 Canceling the Comparator Function HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Page 126: ...112 Excel Direct Entry Function HID Connection HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Page 184: ...170 Index HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Page 185: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Page 186: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...