121
Basic Specifications
9.2
Basic Specifications
Measurement
parameters
• Measuring battery internal resistance
• Measuring battery terminal voltage (DC voltage only)
• Measuring temperature
Measurable
ranges
Resistance measurement:
0.000 m
Ω
to 3.100
Ω
(4-range configuration)
Voltage measurement:
0.000 V to ±60.00 V (2-range configuration)
Temperature measurement:
−10.0°C to 60.0°C (single-range configuration)
Maximum input
voltage
60 V DC (between the positive and negative
measurement terminals)
AC voltage cannot input.
Maximum rated
voltage to earth
60 V DC (Measurement category not rated)
Anticipated transient overvoltage 330 V (between all
measurement terminals and ground)
Measurement
method
Resistance measurement:
AC four-terminal method
Open-circuited terminal voltage:
5 V peak at a maximum
Measurement current:
1.6 mA to 160 mA (Fixed according to the
resistance measurement range setting)
Temperature measurement:
Platinum temperature sensor (500
Ω
at 25°C)
A/D conversion method:
Delta-sigma modulation
Display update rate:
3 times/s (Resistance, voltage, and temperature
measured as a set)
HIOKI BT3554F961-00
HIOKI BT3554F961-00
Summary of Contents for BT3554-50
Page 2: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Page 14: ...vi Contents HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Page 94: ...80 Canceling the Comparator Function HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Page 126: ...112 Excel Direct Entry Function HID Connection HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Page 184: ...170 Index HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Page 185: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...
Page 186: ...HIOKI BT3554F961 00 HIOKI BT3554F961 00...