Setting Threshold Values for the Comparator
4.3 Setting Threshold Values for the
Comparator
This section describes how tot set threshold values for the
comparator (resistance warning threshold, resistance rejection
value, voltage warning threshold).
You can set the threshold values through GENNECT ONE
or GENNECT Cross.
See GENNECT ONE (contained in the accompanying CD)
or User Manual on GENNECT Cross.
Example: Threshold values for a battery that initially has
an internal resistance of 0.4
Ω
and generates
a voltage of 2 V.
Resistance warning threshold: 0.6
Ω
(1.5 times the initial value)
Resistance rejection value:
0.8
Ω
(2 times the initial value)
Voltage warning threshold:
1.8 V
*: The initial values mean the resistance value a fresh battery or
non-defective battery has and the initial voltage value such a
battery can generate.