51
Adjusting the Zero Point (Zero Adjustment)
How to short-circuit various test lead
For Pin Type Lead
Use the included or optional zero-adjustment board. The zero
adjustment can be achieved based on the AC four-terminal method.
1
Choose two holes on the zero-
adjustment board whose
interval is nearly equal to the
distance between the two
terminals on a battery to be
measured.
2
Press the test lead against the zero-adjustment board in a
vertical direction.
Insert the pins with the marked
(engraved) side of the probe facing you.
Model 9465-10 or
Model L2020 Pin Type Lead
Model 9772 Pin Type Lead
HIOKI BT3554F961-00
HIOKI BT3554F961-00