5.4 Commands Specific to the 3157
────────────────────────────────────────────────────
:TIMer
■
Enables and disables the test time.
:TIMer?
■
Queries the test time enablement.
:TRANsmit:TERMinator
■
Sets the data terminator for response messages.
Syntax
:TIMer
_<data>
<data>
ON/OFF (character data)
Function
Enables and disables the test time.
Note
When the endless timer function is set, the ON/OFF setting for the test time
is not reflected in the test.
Example
Transmission
:TIMer ON
The test time is set to ON.
Error
If <data> is other than character data and numerical value described above, a
command error occurs.
The execution of this command in a state other than the following states
causes an execution error:
・
READY state
・
Optional function setting screen
Syntax
:TIMer?
Function
Returns the current enablement state of the test time as character data.
<data>
ON/OFF (character data)
Example
Response
If headers are on
:TIMER ON
If headers are off
ON
Error
If the response message is longer than 300 bytes, a query error is generated.
Syntax
:TRANsmit:TERMinator
_<data>
<data>
Numerical data in NR1 format between 0 and 255
Function
Sets the data terminator for response messages.
The numerical value can be in NRf format, but any digits after the decimal
point will be rounded.
If <data> = 0, the terminator is set to LF (line feed) + EOI signal.
If <data> = 1 through 255, the terminator is set to CR (carriage return) + LF
and EOI signal.
When powering on, <data> is initially set to 0 (LF+EOI).
Example
Transmission
:TRANsmit:TERMinator 0
The data terminator is set to LF + EOI.
Error
If <data> is other than numerical value described above, an execution error
occurs.