5.4 Commands Specific to the 3157
────────────────────────────────────────────────────
:CONFigure:VUPPer
■
Sets the maximum test value (voltage).
:CONFigure:VUPPer?
■
Queries the maximum test value (voltage).
Syntax
:CONFigure:VUPPer
_<data>
<data>
Numerical data in NR2 format between 0.00 and 6.00
Function
Sets the maximum test value (voltage).
The numerical value can be in NRf format, but rounding is performed for
figures beyond the last valid decimal place.
Note
The maximum test value (voltage) can be used as the basis for the test only
when the maximum and minimum test values are set to be expressed in
"VOLT" (voltage) and when "ON" is selected in the ON/OFF setting for the
maximum test value.
Example
Transmission
:CONFigure:VUPPer 2.50
The maximum test value (voltage) is set to 2.50 V.
Error
If <data> is other than numerical value described above, an execution error
occurs.
The execution of this command in a state other than the READY state causes
an execution error.
Syntax
:CONFigure:VUPPer?
Function
Returns the maximum test value (voltage) setting as a numerical value in
NR2 format.
<data>
Numerical data in NR2 format between 0.00 and 6.00
Example
Response
If headers are on
:CONFIGURE:VUPPER 2.50
If headers are off
2.50
Error
If the response message is longer than 300 bytes, a query error is generated.