5.4 Commands Specific to the 3157
────────────────────────────────────────────────────
:SYSTem:OPTion:MOMentary
■
Sets the momentary OUT function.
:SYSTem:OPTion:MOMentary?
■
Queries the momentary OUT function.
Syntax
:SYSTem:OPTion:MOMentary
_<data>
<data>
Numerical data in NR1 format between 0 and 1
Function
Sets the momentary OUT function.
0: Not set
1: Set
The numerical value can be in NRf format, but any digits after the decimal
point will be rounded.
Note
The ":STARt" command works in a state in which the momentary OUT
settings are disabled even though the momentary OUT settings have been set.
Example
Transmission
:SYSTem:OPTion:MOMentary 1
The momentary OUT function is set.
Error
If <data> is other than numerical value described above, an execution error
occurs.
In the continuous test mode, setting the momentary OUT function causes an
execution error.
The execution of this command in a state other than the following states
causes an execution error:
・
READY state
・
Optional function setting screen
Syntax
:SYSTem:OPTion:MOMentary?
Function
Returns the momentary OUT function setting as a numerical value in NR1
format.
<data>
Numerical data in NR1 format between 0 and 1
Example
Response
If headers are on
:SYSTEM:OPTION:MOMENTARY 1
If headers are off
1
Error
If the response message is longer than 300 bytes, a query error is generated.