2.5 Continuity Check
索引
4
3
2
1
7
6
5
10
9
8
2.5 Continuity Check
Never apply voltage to the test leads. Doing
so may damage the instrument and result in
personal injury. To avoid electrical accidents,
remove power from the circuit before mea-
suring.
1.
Set the function switch.
2.
Select continuity (
) with
the
SHIFT
key.
3.
If you want to change the
range, press the
RANGE
key.
(Normally, manual range is
used.)
To change to auto range, hold
the
RANGE
key down.
4.
Connect the test leads to
the test terminals.
5.
Connect the test leads to
the object being tested.
6.
Read the value displayed.
In each range, the buzzer
sounds when the count is less
than 100 (less than 10.0
Ω
in the
999.9
Ω
range).
However, in relative value dis-
play mode, this conforms with
the internally measured value,
not the display value.
<Example>
(
AUTO
off)
Black
Red
Red
Black
(
lights)