42
―――――――――――――――――――――――――――
Chapter 3
Specifications
――――――――――――――――――――――――
Measurement
conditions
Singe phase/balanced three phases,
50/60 Hz, sine wave
Measurement range
Effective measurement current range:
1 A to 1000 A
Effective measurement voltage range:
80 V to 600 V
Phase angle measurement
φ
Method of
measurement
Measurement range
Obtained from phase detection circuit.
Measurement
Mode
Resolution
Measurement Range
Accuracy
φ
0.1
LEAD 90
to 0 to LAG 90
3
Power factor measurement
λ
Method of
measurement
Measurement range
Obtained by calculation from phase
angles.
Measurement
Mode
Resolution
Measurement Range
Accuracy*
cos
φ
0.001
LEAD 0 to 1 to LAG 0
3
2dgt.
* Calculating error of
2 dgt. is added to phase angle measurement error.
Reactive factor measurement
Method of
measurement
Measurement range
Obtained by calculation from phase
angles.
Measurement
Mode
Resolution
Measurement Range
Accuracy*
sin
φ
0.001
LEAD 0 to 1 to LAG 0
3
2dgt.
* Calculating error of
2 dgt. is added to phase angle measurement error.
3.1.4 Specifications of Power Factor and Phase
Angle Measurements 1
φ
PF Meter and 3
φ
PF Meter