19
―――――――――――――――――――――――――――
Chapter 2
Measurement Procedure
――――――――――――――――――――――――
P
o
w
e
r
s
u
ppl
y
side
Lo
ad
Figure 6. Current measurement
Red
Yellow Black
Lo
ad
P
o
w
e
r
s
u
ppl
y
side
Figure 7. Voltage measurement
[Current measurement]
When only measuring current, the orientation of the
clamp sensor is irrelevant. Moreover, the voltage
cord need not be connected to the instrument.
[Voltage measurement]
When only measuring voltage, the clamp sensor
need not be clamped.