35
―――――――――――――――――――――――――――
Chapter 2
Measurement Procedure
――――――――――――――――――――――――
Display 1
Item No.
Display 2
Item Name
Display 3
Setting
Initial
Value
2.9 SETUP Function
The settings for this instrument are made in SETUP
mode.
In SETUP mode, you can make settings for
measurements, display, and ancillary functions.
1. Hold down the
SET1
key while powering on the
instrument by pressing the
POWER
key. This
activates SETUP mode.
2. Select a setting item. The
MODE
key increments
the item No.; the
MAX/MIN
key decrements the
item No.
3. The settings can be modified using the
U/
key or
I/
key.
4. Pressing the
HOLD
key twice in succession restores
the initial values for the setting items.
5. At instrument power-off, "
" (SAVE END)
appears and the settings are saved.
6. Details of Settings
(1) Setup of single-phase power meter system
Item No. 1-01
ON Sets 1
φ
PF meter.
OFF Sets 1
φ
P meter.