51
_______________________________________________________
4.1 Measurement Specifications
_______________________________________________________
Range
(Accuracy Range)
Resolution
10 Hz (1.00 to 10.00 Hz)
0.01 Hz
0.3%rdg. 1dgt.
100 Hz (10.0 to 100.0 Hz)
0.1 Hz
0.3%rdg. 1dgt.
1000 Hz (100 to 1000 Hz)
1 Hz
1.0%rdg. 1dgt.
Current Specifications
Maximum input
current
2000 Arms continuous, 2840 Amax.
See Fig.3
Effect of conductor
position
within
0.7% (in any direction from
sensor center)
External magnetic
field interference
AC 400 A/m (external magnetic fields)
corresponds to 2 A or less (display)
Maximum rated
voltage to earth
max. 600 Vrms
Fig.3 Frequency-dependent deletion characteristics