37
_______________________________________________________
2.6 Continuity Check
_______________________________________________________
NOTE
5. Then connect the test leads to the test points for
measurement.
The beeper sounds when the measured resistance is
below about 30
Ω.
・
If voltage is applied inadvertently, the internal
circuitry is protected up to 600 VAC.
・
The measured resistance value appears on the digital
display.
・
If voltage larger than approximately 2 or 3 V is
input, the Continuity check mode cannot be
selected.
・
If the display shows larger than 7 or 8 A at 200 A
range in the DCA mode, the Resistance
measurement mode cannot be selected. Press the
0ADJ/RESET
key to reset the display to 0.
Summary of Contents for 3285-20