34
_______________________________________________________
2.5 Resistance Measurement
_______________________________________________________
DANGER
Never apply voltage to the test leads when the
Resistance measurement mode is selected. Doing
so may damage the instrument and result in
personal injury.
To avoid electrical accidents, remove power from
the circuit before measuring.
CAUTION
For safety reasons, when taking measurements, only
use the L9207-10 Test Lead provided with the
instrument.
2.5 Resistance Measurement
1. Press
Ω
/
button to display
Ω
. Auto-ranging is
used for resistance measurement.
2. Insert the red test lead to
V/
Ω
and the black test
lead to
COM
of the voltage/resistance terminal.
3. Attach or remove the rigid insulating sleeve as
required by the measurement object.
4. Touch the tips of the red and black test leads
together and verify that 0
Ω
is displayed.
5. Then connect the test leads to the test points for
measurement.