46
_______________________________________________________
3.4 Storing Values in Memory (Recording Function REC)
_______________________________________________________
NOTE
4. Resetting of Recording Function
Push
0ADJ/RESET
button, in the case that data is
reset during the recording function action.
5. Cancellation of Recording Function
To cancel the recording function, press the related
function button (
or
) for the measurement
in progress. Once the recording function is
canceled, the auto power-off function becomes
effective. (
APS
goes on.)
For a long term measurement, check how much the
battery power remains. (Shown on the bar graph at
power on.)
When starting the recording function (
REC
) in an
auto range, the range is set as the range of when
that pushed
MAX/MIN
button.
When you need minimum value and average value
data, make sure to activate the recording function
during measurement. If the function is activated
when there is no input, the minimum value will
remain zero. To deactivate the recording function,
press the
HOLD
button and read minimum value
and average value to terminate measurement. If you
disconnect the clamp or test lead from the circuit
under measurement without deactivating the
recording function beforehand, the minimum value
will be zero.
When the instrument is turned off, accumulated data
are lost.
Summary of Contents for 3285-20
Page 5: ......