46
The Vest® Airway Clearance System, Model 105
Instructions for Use
(145330 REV 18)
Use The Vest® Airway Clearance System, Software Version 1.22 or 2.00 and higher
USE THE VEST® AIRWAY CLEARANCE SYSTEM,
SOFTWARE VERSION 1.22 OR 2.00 AND HIGHER
Make sure the unit is set up
correctly. Once power is
connected, the unit will show
a blank screen for up to 15
seconds. Then it will show the
model number and the total
therapy hours for the unit.
This screen shows for 10 seconds or until you press the
ON
button.
NOTE:
To show the total therapy
hours, press the
OFF
button
when the
Main
menu is
shown.
N
ORMAL
M
ODE
Normal Mode is using one Frequency, Pressure, and Time
settings for the entire therapy session.
1.
Press the
Up
arrow
button above
NORMAL
for the standard
operation mode. The
screen will show the
previously programmed
Frequency, Pressure, and
treatment Time settings (either factory settings or settings
programmed by your clinical trainer based on your doctors
prescription).
2.
Confirm the settings on the screen. If the settings match the
prescribed treatment, continue to step 3. Otherwise, adjust the
settings to match the prescription.
a.
To adjust the
Frequency setting,
press the left
Up
or
Down
arrow buttons
until the prescribed
treatment shows on
the screen. The
Summary of Contents for Vest 105