The Vest® Airway Clearance System, Model 105
63
Instructions for Use
(145330 REV 18)
Use The Vest® Airway Clearance System, Software Version 1.22 or 2.00 and higher
5.
Press the right
Up
arrow
button above
pgm
disable
(it will turn to
PGM DISABLE
).
6.
To accept the change,
press the right
Down
arrow below
ACCEPT
.
7.
To accept all changes and
exit
out of the menu,
press the right
Down
arrow button below
ACCEPT
.
8.
Once you have accepted the change, the system will automatically re-
boot and go to the main screen of
NORMAL
Mode. You will no longer
be able to see the Program or Ramp mode programs.
9.
To switch back to Program Mode, repeat the steps starting at step 3.
NOTE:
Press the
Up
arrow button above
pgm enable
at step 4 to enable the
Program Mode (it will change
pgm enable
to
PGM ENABLE
).
S
ET
C
USTOM
D
EFAULT
S
ETTINGS
Allows the user to set the default settings for the modes of operation.
These settings will be the initial settings for all users.
1.
Press the
ON
button.
2.
Press the
Up
arrow button
above
NORMAL
. The
Normal mode screen
shows.