The Vest® Airway Clearance System, Model 105
Instructions for Use
(145330 REV 18)
Use The Vest® Airway Clearance System, Software Version 1.22 or 2.00 and higher
4.
Look at the settings on
the screen.
Ramp Time
Percent
is the portion of
the treatment Time
required to increase from
the initial settings (
Ramp
Start
) to the final settings
(
Ramp End
). To adjust the percent to the prescribed therapy, press
the right
Up
or
Down
arrow buttons.
5.
Press the
ON
button to advance to the next screen.
6.
Review the settings for Frequency and Pressure for the initial setting,
Ramp Start
.
a.
To adjust the
Frequency setting,
press the left
Up
or
Down
arrow buttons
until the prescribed
treatment shows on
the screen. The
Frequency (or Hertz (Hz)) may be set between 5 and 20 cycles per
second.
b.
To adjust the Pressure
setting, press the
middle
Up
or
Down
arrow buttons until
the prescribed
treatment Pressure
shows on the screen.
Pressure may be set between 1 and 10.
7.
Press the
ON
button to advance to the next screen.
8.
Review the settings on the screen. Frequency and Pressure for the
final setting,
Ramp End
, are shown as treatment Time.
a.
To adjust the Frequency setting, press the left
Up
or
Down
arrow
buttons until the prescribed setting shows on the screen.
Frequency or Hertz (Hz) may be set between the Ramp Start
setting and 20 cycles per second.
b.
To adjust the Pressure setting, press the middle
Up
or
Down
arrow buttons until the prescribed setting shows on the screen.
Pressure may be set between the Ramp Start setting and 10.