The Vest® Airway Clearance System, Model 105
81
Instructions for Use
(145330 REV 18)
Specifications
Recommended Separation Distances Between Portable and Mobile
RF Communications Equipment and The Vest® Airway Clearance
System, Model 105
The Vest® Airway Clearance System, Model 105 is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or user of
The Vest® Airway Clearance System, Model 105 can help prevent electromagnetic interfer-
ence by maintaining a the minimum distance between portable and mobile RF communi-
cations equipment (transmitters) and The Vest® Airway Clearance System, Model 105 as
recommended below, according to the maximum output power of the communications
equipment.
Rated Maximum
Output Power of
the Transmitter
Watts (W)
Separation Distance According to the Frequency of the
Transmitter Meters (m)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,7 GHz
0,01 W
0,12 m
0,04 m
0,07 m
0,1 W
0,37 m
0,11 m
0,22 m
1 W
1,17 m
0,35 m
0,70 m
10 W
3,69 m
1,11 m
2,21 m
100 W
11,67 m
3,50 m
7,00 m
For transmitters rated at a maximum output power not listed above, the recommended
separation distance
d
in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where
P
is the maximum output power rating of the trans-
mitter in watts (W) according to the transmitter manufacturer.
NOTES:
•
At 80 MHz to 800 MHz, the separation distance for the higher frequency range
applies.
•
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
d
3.5
3
-------
P
=
d
3.5
10
-------
P
=
d
7
10
------
P
=