Caregiver Controls
26
Bed Scale (Optional)
The Advanta™ Bed scale has an accuracy of 1%, and an
operating range of 0 to 450 lb (0 to 204 kg). The scale
display and six scale keys are located on the footboard
control panel. For best results using the scale, follow the
instructions carefully.
Before Operating The Scale
The scale is very sensitive. The weight reading will be most accurate if the bed is not touching anything. This includes
the headwall, lines such as pendant controls, ventilators, or drainage bags. Anything that affects the weight on the bed
even slightly will cause an incorrect weight to appear on the display.
It is also important that no one touch the bed while the scale is operating. For example, touching the footboard can add
several pounds to the displayed weight.
Before placing the patient on the bed, prepare the sleep surface to accurately weigh the patient. Place all linens,
blankets, pillows, or other items the patient will be using on the bed. A list of these items posted near the bed could be
helpful for future reference.
Turning the Scale On or Off
If the scale display reads “OFF,” press the On/Off key to turn on the scale power.
NOTE:
The scale automatically turns off after 5 minutes.
NOTE:
Be sure to place any items on the bed that the patient will be using (linens, pillows, blankets, etc.).
Zeroing the Scale
The bed scale must be set at zero before weighing a patient.
•
Set the scale at zero by pressing and holding the Zero/000.0 key on the scale control panel.
•
Continue holding the Zero/000.0 key until the scale display “000.0” stops flashing; then release the key.
•
Important! After you release the Zero/000.0 key, the scale display “CALC” will appear. Do not touch the bed
until the display stops flashing “CALC” and a beep sounds. This may take up to 20 seconds.
•
Place the patient on the bed. For the most accurate reading, the patient must be lying still and centered on the
mattress.
Changing Items on the Bed
The Advanta™ Bed scale can compensate for items that are added to or removed from the bed. Use the
scale control panel to change (add or remove) items on the bed.
•
Press and hold the Change Items key on the scale control panel until the displayed weight stops flashing; then
release the key. The scale holds the patient’s weight in the memory while items are being changed.
•
Change the items on the bed.
Advanta™ Bed Scale Controls