45
2.9 PV
:
Measurement
⚝ᒁ
4
3
2
1
7
6
5
10
9
8
• Insulation resistance is a ratio of a voltage applied and
a leakage current. Indication values are not sometimes
stable depending on the object to be measured but it is
not a failure.
• The
MEASURE
key should be fully pressed until the
hot-line warning indication lights. Insufficient press
may result in an incorrect measurement.
• After using this instrument, turn OFF the rotary switch.
• When you test an electrical line connected to
equipment having a withstand voltage lower than the
test voltage or equipment/part with an unknown
withstand voltage, it is recommended to make any
measurement after they are disconnected from the
line.
• As solar panel to ground capacitance is larger, it may
take time until measured values become stable.
• The PV
:
measurement function does not comply with
EN61557 requirements. To perform measurements in
conformity with EN61557, select the insulation
resistance range.
• If open voltage of solar cell string is higher than test
voltage, accurate measurement cannot be achieved.
PV
:
500 V range should be used with an open voltage
500 V or lower and PV
:
1000 V range for an open
voltage 1000 V or lower.
• If a voltage higher than test voltage is being generated,
no measurement is available.
• As PV
:
measurement uses a current-limiting resistor
of 1 M
:
connected to EARTH terminal, the output
voltage is divided by the 1 M
:
resistor and the resistor
connected between the measuring terminals. For
example, when a 10M
:
resistor is used, the output
voltage is divided by 1 M
:
and 10 M
:
.