22
2.2 Pre-measurement inspection
2.2
Pre-measurement inspection
No
Exposed
Do not use if
damage is pres-
ent as you could
receive an elec-
tric shock.
Contact your
authorized Hioki
distributor or
reseller for
replacements.
Isn’t the battery exhausted?
Set the rotary selector away
from OFF and confirm the
battery indicator.
Replace the
batteries.
See:
(p.68)
Isn't the white portion (insu-
lation layer) inside the
cable exposed?
Checking the remaining battery life
Checking the test lead
No Exposed
1. Use the rotary selector to
select Insulation Resis-
tance.
2. Short the test lead tips.
3. Press the
MEASURE
key.
Does the display show
0 M
:
?
Yes
Inspection complete
Please read the “Operating Precautions” (p.9) before use.
The battery indicator
is flashing.
There is battery life remaining.
The following
issues may be
occurring:
• The test lead has
not been inserted
all the way.
o
Insert the test
lead all the way.
• There is a broken
connection in the
test lead.
o
Replace with
the designated
test lead.
www.
.com