4.3 Replacing the Pin (Option)
The pin at the front of the Model L9788-10 (option) can be
replaced when it has worn away or is damaged. Replacement
pins are available at your authorized Hioki distributor or reseller.
(Model L9788-90 Tip Pin)
• To clean the instrument, wipe it gently with a soft cloth
moistened with water or mild detergent. Never use solvents
such as benzene, alcohol, acetone, ether, ketones, thinners or
gasoline, as they can deform and discolor the case.
• Wipe the LCD gently with a soft, dry cloth.
• As dirt accumulated on the metal part of alligator clip has an
effect on measurements, keep the surfaces clean by gently
wiping with a soft cloth.
4.3 Replacing the Pin (Option)
1.
Turn off the power of the instrument and disconnect the
L9788-10.
2.
Rotate the tip pin with a spanner (7 mm width) to remove it.
3.
Turn the new tip pin with a spanner and attach it to the
L9788-10. (tightening torque: 0.3N•m)
4.
Check the performance. Measure an object with a known
resistance. Make sure that the measured resistance is
correct before using the L9788-10.
Turn to remove
4.4 Cleaning
www.
.com
1.800.561.8187