DTS-60A / DTS-100A O
PERATORS
M
ANUAL
8.3.2 test read <test ID#>
The
test read <test ID#>
command displays information about the selected test id#.
The first block of information table includes: Test Name; Electrode Spacing and Type;
Number of Stages; Start Up Wait Timer; and Start Up Stirrer Timer (see section 4.1).
The second information block includes information about individual test stages, such as:
Stage ID#; Voltage Rate or Settings; Voltage Units; Duration; Stirrer Timer; After Breakdown Wait
Timer; Stirrer Timer; After Breakdown Stage GoTo; and finally the Breakdown Voltage Storage On/Off.
Example of a Standard Test:
DTS60A > test read 2
----------------------------------------------------------------------------------------------
Test Name : ASTM-D1816/84
Electrode Spacing : 0.04 +/- 0.0010in
Electrode Type : Type II
Number of Stages : 05
Start-Up Wait Time : 00:03:00
Stirrer During Start-UP : 00:03:00
----------------------------------------------------------------------------------------------
| Stage | Voltage | Duration | Stirrer | WaitTimer | Stirrer | GoTo Stage | BD Store |
----------------------------------------------------------------------------------------------
| 00 | 000.5 kV/s | breakdown | 00:10:00 | 00:01:00 | 00:01:00 | next | yes |
| 01 | 000.5 kV/s | breakdown | 00:10:00 | 00:01:00 | 00:01:00 | next | yes |
| 02 | 000.5 kV/s | breakdown | 00:10:00 | 00:01:00 | 00:01:00 | next | yes |
| 03 | 000.5 kV/s | breakdown | 00:10:00 | 00:01:00 | 00:01:00 | next | yes |
| 04 | 000.5 kV/s | breakdown | 00:10:00 | 00:01:00 | 00:01:00 | stop | yes |
----------------------------------------------------------------------------------------------
DTS60A >
Example of a User Defined Test:
DTS60A > test read 13
----------------------------------------------------------------------------------------------
Test Name : Eval-mini
Electrode Spacing : 2.5 +/- 0.250mm
Electrode Type : Type I
Number of Stages : 03
Start-Up Wait Time : 00:00:05
Stirrer During Start-UP : 00:00:05
----------------------------------------------------------------------------------------------
| Stage | Voltage | Duration | Stirrer | WaitTimer | Stirrer | GoTo Stage | BD Store |
----------------------------------------------------------------------------------------------
| 00 | 003.0 kV/s | breakdown | 00:00:05 | 00:00:05 | 00:00:05 | next | yes |
| 01 | 003.0 kV/s | breakdown | 00:00:05 | 00:00:05 | 00:00:05 | next | yes |
| 02 | 003.0 kV/s | breakdown | 00:00:05 | 00:00:05 | 00:00:05 | stop | yes |
----------------------------------------------------------------------------------------------
DTS60A >
68