8.2.6 contrast Command
The
contrast
command is used for asking or setting the LCD contrast, Figure 8-7. Possible values are
0 to 15.
If no argument is given, the system returns the current contrast setting value.
DTS60A > contrast
Current Contrast Value: 07
DTS60A > contrast set 10
New Contrast Value: 10
DTS60A >
8.3 Test Commands
The
test
command is used to list the Standard Tests and User Defined Tests contained in the system.
This command also has the ability to list the individual stages of a given test. There is also the option
of deleting a User Defined Test in case of error, as well as a command to clear all User Defined Tests.
This is also the command which allow creation of new user defined tests.
8.3.1 test list
The
test list
command lists all tests present in the system, showing their test ID # which is needed
for the read and delete functions.
For all tests, the number of stages is shown.
For User Defined Tests, the Time & Date when the test was created is shown.
DTS60A > test list
Available Tests List...
--------------------------------------------------
ID : Name Creation Time & Date Stages
--------------------------------------------------
00 : IEC-156/95 - 06
01 : IEC-156/63 - 06
02 : ASTM-D1816/84 - 05
03 : ASTM-D877/87 - 05
04 : UTE-C27221 - 06
05 : CEI-10-1/73 - 06
06 : VDE-0370 - 06
07 : IRAM-2341/72 - 06
08 : RVHP-1985 - 06
09 : PN-77 - 06
10 : Evaluation 20:34:02 09/20/2016 30
11 : Eval_400v 20:38:39 09/20/2016 30
12 : Eval_100v 20:39:42 09/20/2016 30
13 : Eval-mini 07:59:00 10/30/2016 03
--------------------------------------------------