DTS-60A / DTS-100A O
PERATORS
M
ANUAL
48.2
kV
STIR
0.0
kV
STIR
STEP 5
BACK
By pressing the BACK key the STIRRER is
enabled or disabled. While enabled the system
displays the STIRRER ICON.
0.0
kV
STIR
STEP 6
ENTER
Press the ENTER key to start the test.
1.0
kV
STIR
STEP 7
The High Voltage On indicator is depicted.
Voltage is increasing at the rate of 0.5kV/s
and the stage has been running for two seconds.
The voltage measurement in kV is updated
continuously.
STEP 8
Breakdown Detection occured!
The Breakdown icon is blinking.
The High Voltage On indicator is off.
The voltage at which breakdown occurred is shown.
STAGE
1
STAGE
1
STAGE
1
HN on
STAGE
1
TEST:SIMPLE TEST
VOLTAGE:
0.5 kV/s
TEST:SIMPLE TEST
VOLTAGE:
0.5 kV/s
TEST:SIMPLE TEST
VOLTAGE:
0.5 kV/s
TEST:SIMPLE TEST
VOLTAGE:
0.5 kV/s
TIMER:00:00:00
TIMER:00:00:00
TIMER:00:00:02
TIMER:00:01:36
12:10 01/08/16 65°F
12:10 01/08/16 65°F
12:10 01/08/16 65°F
PRESS ANY KEY...
Figure 3-8b –
Simple Test Example
33