![High Voltage DTS-100A Operator'S Manual Download Page 67](http://html.mh-extra.com/html/high-voltage/dts-100a/dts-100a_operators-manual_2130380067.webp)
DTS-60A / DTS-100A O
PERATORS
M
ANUAL
7.3 Printout Description
Upon completion of a test, the user is prompted to print the results of the test. In Fig. 7-7 is depicted
the printed information.
1. High-Voltage Logo
2. Operator information
3. Time and Date
4. Environmental temperature
5. Test Name
6. Electrodes
7. Number of Breakdowns
8. Mean Value of Breakdown Voltages
9. Standard Deviation of Breakdown Voltages
10. Stages/ Starting Voltage/ Ending Voltage/ Duration/ Breakdown
----------------------------------------
HighVoltage, INC.
DTS-60A Test Report
Version 1.00
----------------------------------------
Operator
Company : HighVoltage
Name : John Smith
Contact : +30-210-9319505
Info : [email protected]
----------------------------------------
Report ID : 01018
----------------------------------------
Time : 14:41
Date : 12/05/2016
Temperature : 70°F
Test Name : ASTM-D1816/84
Electrodes : Mushroom electrodes
Num. Breakdown : 5
Mean Breakdown : 50.61
Std Breakdown : 6.24
Stage 1: 0.5kV/s, 0.0> 48.1kV, Breakdown
Stage 2: 0.5kV/s, 0.0> 40.7kV, Breakdown
Stage 3: 0.5kV/s, 0.0> 54.9kV, Breakdown
Stage 4: 0.5kV/s, 0.0> 54.3kV, Breakdown
Stage 5: 0.5kV/s, 0.0> 55.1kV, Breakdown
----------------------------------------
Figure 7-7 –
Printout Receipt
61
Summary of Contents for DTS-100A
Page 92: ......