HEBU reusable APC probe
30
EN
9.1 Packaging
Compliant packaging of products for sterilization in line with ISO 11607. Packaging used must be
suitable for the instruments and protect them from microbiological contamination during storage.
The seal must not be under tension. HEBUmedical recommends container or hospital common
sterilization paper/film packagings for sterilized packaging.
10 Lifetime
The steam sterilization procedure was validated by laboratory tests. The products were sterile
validated at a pre-vacuum of at least 5min duration and a temperature of 134°C for a lifetime of 50
cycles.
You can continue to use the instruments at your own responsibility over this cycle value if the
tests described in chapter 7 have been successfully completed.
11 Storage
Storage of the sterilized instruments in a dry, clean and
dust-free environment. Products must be protected from direct
sunlight
in storage.
12 Warranty/Repair
Our products are made of high-quality materials and are carefully checked before delivery. However,
even when used as intended, they are subject to wear to a greater or lesser extent depending on the
intensity of use.
This wear is due to technical reasons and is unavoidable.
However, if faults occur that are not related to wear, please contact our customer service department.
Defective products may no longer be used and must have undergone the entire preparatory process
before being returned.
The flexible APC probes, reusable must not be sterilized in hot air sterilizers.
During validation the instruments were packaged in hospital common sterilization packagings
(paper/film packagings) and steam sterilized.