HEBU reusable APC probe
27
EN
Observe the following by loading:
◼
Place the dismantled/opened instruments securely in the tray.
◼
Products with openings and gaps have to be faced down with the opened side so that they
can be cleaned and no water of the cleaning process is collecting inside them. If available use
balanced devices for rinsing
◼
Do not overload trays, avoid creating any overlaps.
Preliminary rinsing (cold, if applicable fully demineralized water without additives) is followed by
chemical. The chemical cleaning should take place at
40°C -60°C
for at
least 5 minutes
.
We recommend products with a
pH-value within 9-10,
e.g. Neodisher MediClean forte from Dr.
Weigert. The cleaning agents used should be selected depending on the material and properties of
the instruments and in accordance to national regulations: If there is a high chloride concentration in
the water, pitting and tension crack corrosion can occur on the products. The occurrence of this type
of corrosion is minimized by using alkaline cleaning agents and demineralized water. By adding an
acid-based neutralization medium, the rinsing off of alkaline cleaning agent residues is facilitated
during the first intermediate rinsing process (warm or cold water).
In order to prevent the formation of deposits, it is advisable to use neutral cleaners where the water
quality is unfavourable. After the second intermediate rinsing process, thermal disinfection takes
place.
The thermal disinfection should take place at temperatures of between
80 and 95°C
, with an
exposure time as outlined in EN ISO 15883
.