HEBU reusable APC probe
28
EN
After the finishing of the program take the good out of the machine because corrosion can arise if the
instrument remains in the machine.
Parameters used for the validation of preparation
Pre-rinsing
1 minute with cold tap water
Cleaning
Temperature: 55 °C
Soaking Time: 5 minutes (worst case)
Neodischer Mediclean forte 0,4% (worst case)
Neutralization
Temperature: cold DI water
Soaking Time: 2 minutes
Neodisher Z 0,1%
Post-rinsing
2 minutes with cold DI water
Disinfection
Temperature: 90 °C (A
0
3000)
Soaking Time: 5 minutes
7.7 Drying
Ensure adequate drying by the cleaning and disinfection device or using other suitable measures.
8 Maintenance, inspection
After cooling to room temperature, the products must be visually inspected for protein residues and
other contamination, paying particular attention to cavities, blocks, inclusions, pipes, and other
inaccessible areas. Instruments which are not free of residues must be returned for a complete
retreatment process.
Carry out the functional check mentioned above. Products with stains, which are blunt, bent, no
longer function or which are otherwise damaged must be segregated!
To help identify faulty products that need to be sorted out, we recommend the brochure "Instrument
Reprocessing" from the Working Group “Instrumenten Aufbereitung”. This includes in particular
Chapter 8 "
Checks and Care" and Chapter 12 "
Surface Changes: Deposits, Discoloration,
Corrosion, Aging, Swelling and Stress Cracks".
Drying was omitted in the validation (worst case condition).