![HEBU medical HF 9928-01 Operating Manual Download Page 22](http://html1.mh-extra.com/html/hebu-medical/hf-9928-01/hf-9928-01_operating-manual_2118150022.webp)
HEBU reusable APC probe
22
EN
5 Danger notice
The products may only be used in combination with HEBU accessories and only by clinically trained
personnel. Electrodes must not be bent, as this may damage the material and cause the electrode to
break. The output power of the HF device may only be set to the absolutely necessary value for the
intervention. If the usual coagulation power is not achieved despite the standard settings of the HF
unit, never increase the output power of the unit without first checking it. Check in detail Proper
contact of all HF plugs and cables, functioning of the foot switch or finger switch on the handle,
insulation of the HF cables, the instrument, and cleanliness and wear of the distal end of the active
electrode(s). Before starting the operation, check that the monitors are transmitting signals without
interference (e.g. without noise) by pressing the yellow key (CUT) and the blue key (COAG).
To pull the HF cable, always touch the plug.
Never pull on the HF cable
, as this could result in
damage. Cables with defective/brittle insulation or a break in the electrical cable (due to severe
kinking or crushing of the cable) can cause burns to the user/patient or even cause fire. Active HF
instruments and HF handles must not be placed on the patient to prevent personal injury in the event
of accidental HF activation.
To clean the active electrodes, the "Auto Start" mode (only for bipolar application) must be
deactivated or the instrument must be disconnected from the generator. The tip area may
still be hot immediately after activation, i.e. there is a risk of burns.
Endogenous burns are burns caused by high current density in the patient's tissue. Causes may
include: The patient unintentionally receives contact with electrically conductive parts. Capacitive
currents can cause burns when HF cables come into direct contact with the skin.
Exogenous burns are burns caused by the heat of ignited liquids or gases. Explosions are also
possible. Causes may include: ignition of skin cleansers and disinfectants, ignition of anesthetic
gases, etc.
Cardiac pacemakers can be damaged by HF current. Consult a cardiologist before the
procedure. Never perform outpatient procedures with HF current on patients with cardiac
pacemakers.