TEDS transducer identification
T40FH
A4429-3.0
HBM: public
41
Additional templates, such as the
Signal Conditioning
template, can also be
described by the user.
Template: Bridge Sensor
Parameter
1)
Unit
Required
user
rights
Explanation
Transducer Elec
trical Signal Type
Bridge
Sensor
ID
Minimum Torque
0.000
N
@
m
CAL
The physical measurand
and unit are defined
when the template is cre
ated, after which they
cannot be changed.
Maximum Torque
150000
N
@
m
CAL
Minimum Electri
cal
Value
0.0000m
V/V
CAL
The difference between
these values is the sensi
tivity according to the
HBM manufacturing cer
tificate or from the cali
bration.
Maximum Electri
cal Value
1.8245m
V/V
CAL
Mapping Method
Linear
This entry cannot be
changed
Bridge type
Full
ID
The bridge type. "Full" for
a full bridge.
Impedance of
each bridge ele
ment
1550+-100
ohm
ID
Input resistance accord
ing to the HBM data
sheet
Response time
1.0000000u
s
ID
Of no significance to
HBM transducers
Excitation Level
(Nominal)
5.0
V
ID
Nominal (rated) excita
tion voltage according to
the HBM data sheet