Functions and outputs
Quantum
X
I3031-14.0
HBM: public
177
Notice
After an analog output is configured, its function (configu
ration, scaling) continues to be available even if the com
puter is disconnected. Therefore no connection to a PC
is required.
Specifically MX410B supports furthermore 8 peak value
detection channels and 4 RMS channels.
These functions can be used to generate so-called virtual
signals that can also be output at the analog output and
made available to the QuantumX system. This also
makes the signals visible to the software.
The device is parameterized by means of the software
(e.g. QuantumX Assistant or catman®AP).
Keep in mind the following notes when working with peak
value monitoring channels:
S
The maximum output (sampling rate) is restricted to
4800 Hz
S
PEAK values can only be reset via PC software (sys
tem signals will follow)
S
The output rate of the peak value monitoring channel
must not be higher than the sampling rate of the input
channel.
S
Filters set for the MX410B do not apply to peak value
monitoring channels
S
These channels are always unfiltered. However, the
input signal is filtered.
S
The peak value units do not accept other peak value
units or RMS as input. Only the 4 analog inputs are
allowed.