![HBM CX27B Operating Manual Download Page 126](http://html.mh-extra.com/html/hbm/cx27b/cx27b_operating-manual_2111759126.webp)
Modules and transducers
126
I3031-14.0
HBM: public
Quantum
X
CAN LEDs (channel)
Constant green
Channel is ready for operation
Flashing yellow
Firmware1 download active
Yellow on
Boot process running
Red on
Channel has errors
Ethernet LED
Green on
Ethernet link status is OK
Flashing yellow
Ethernet data transmission ongoing
8.9.4
Receiving CAN messages
To be able to receive CAN messages, the node must be
able to identify the relevant messages. This can be done
directly on the node or, in a reproducible way, by previ
ously generated messages in the sensor database. Indi
vidual messages can be linked to the node by dragging
from the sensor database and dropping them where re
quired.
CAN databases type *.dbc can also be read into the
sensor database. If no CAN database is available, it can
also be created. Editors for this purpose are provided by
different companies.
Received CAN messages are instantly "time-stamped" in
measurement mode. This enables directly acquired
measured quantities and CAN messages to be acquired
and analyzed in parallel and synchronously in the entire
system.