Modules and transducers
120
I3031-14.0
HBM: public
Quantum
X
"Calibration Table" if no additional optional templates are
used.
This function delivers the best results when the ambient
temperature of the MX1609/KB/TB, and therefore the
temperature of the cold junction, is kept constant.
Conditions for using RFID chips for measuring point
identification:
S
All channels can read/write via RFID
S
The neighboring channel must not be occupied in the
MX1609/KB/TB during writing
S
Maximum distance chip to housing: 1 mm
S
For self-assembly: Check position of chip on plug
Thermocouple plug with integrated RFID chip from
HBM
RFID chip
The chip for measuring point identification is already in
tegrated in the HBM THERMO-MINI.