![HBM CX27B Operating Manual Download Page 10](http://html.mh-extra.com/html/hbm/cx27b/cx27b_operating-manual_2111759010.webp)
Safety instructions
10
I3031-14.0
HBM: public
Quantum
X
Conditions on site
For modules in the housing with degree of protection
IP20:
S
Protect the modules from dirt and moisture or the ef
fects of weather such as rain, snow, etc.
S
The permissible relative humidity at 31
o
C is 80%
(non-condensing); linear reduction to 50% at 40
o
C.
S
Make sure that the side ventilation openings are not
covered.
For all modules:
S
Do not expose the modules to direct sunlight.
S
Please observe the permissible maximum ambient
temperatures stated in the specifications.
S
Ensure there is adequate ventilation for installation in
the BPX001 backplane.
Maintenance and cleaning
The modules are maintenance-free. Please note the
following points when cleaning the housing:
S
Before cleaning, disconnect all connections.
S
Clean the housing with a soft, slightly damp (not wet!)
cloth.
Never
use solvent as this could damage the
labeling or the housing.
S
When cleaning, ensure that no liquid gets into the
module or connections.
Outputs
Particular attention must be paid to safety when using the
digital, analog or CAN bus outputs of a module. Ensure