![HBM CX27B Operating Manual Download Page 119](http://html.mh-extra.com/html/hbm/cx27b/cx27b_operating-manual_2111759119.webp)
Modules and transducers
Quantum
X
I3031-14.0
HBM: public
119
Standard miniature thermo
couple plug or
1THERMOMINI from HBM
with integrated RFID chip
8.8.1
Thermocouple with TEDS functionality (RFID)
Measuring point identification
An RFID
31)
chip in or on the thermocouple plug ensures
wireless transducer identification through the amplifier.
RFID technology enables contactless reading and writing
of data such as the precise measuring point or the re
quired physical unit (
5
C or
K). Data is entered with the
TEDS Editor provided by HBM. It is then written to the
RFID chip via a corresponding RFID transponder in the
amplifier.
The chip is reusable and works without batteries.
Rescaling
All modules MX1609/KB/TB has a rescaling function. Er
rors from thermocouples or installation situations can be
minimized using a table that converts values from degree
o
C to degree
o
C.
The MX1609/KB/TB can process maximum 64 value
pairs. 14 value pairs can be stored in the TEDS template
RFID = Radio Frequency Identification: Method for communication between transponder and