![HBM CX27B Operating Manual Download Page 151](http://html.mh-extra.com/html/hbm/cx27b/cx27b_operating-manual_2111759151.webp)
Transducer connection
Quantum
X
I3031-14.0
HBM: public
151
9.13 DC voltage sources 10 V
The following amplifiers support a measuring range of
"
10 V
:
MX410B, MX840B, MX440B, MX1601B
Pin assignments for MX1601B
Pin assignment for module MX1615B
9
4
Hsg.
14
(-)
(+)
U
Cable shield
Hsg. = Housing
6
1
1 2 3
1-wire EEPROM (optional)
view from below
1
5
6
10
11
15
2 Data
3 No function
1
MX1615B
6
4
5
7
8
1
1
5
2
8
4
MX1601B
Maximum input voltage
to housing and
supply ground:
MX840B, MX440B, MX1601B:
"
60 V
MX1615B:
"
15 V
MX1601:
"
15 V
Sensor supply:
5 V
…
24 V:
0 V:
Pin 12
Pin 11
Pin 6
Pin 7
1
8
Push in
1
8
Push in
MX1601B: Only channels 1-8 offer an individual sensor
supply of 5-24 V
Channels 9-16 can be activated with a fixed sensor sup
ply (-1 V module supply voltage).