HI 6600 Series Modular Sensor System User Guide
Page | 30
The system can be set-up, calibrated and operated either through the built-in Webserver or through an optional front panel
display. The menu structure above may be followed using either interface.
Using the Webserver
To use the Webserver, a connection must first be established between the Hardy Gateway Module and a connect PC. The
following steps will assist in establishing a connection:
1)
Connect an Ethernet cable between the RJ45 port located on the front of the HI 6600 Gateway Module’s RJ45
and a PC then power up the instrument. A cross-over cable is not required, but can be used.
2)
The instrument will auto-negotiate settings suitable for a variety of operating systems and network
configurations. Wait approximately 15 seconds and check to see if the green light of the Ethernet port is on and
blinking – if so, skip to step 8 – you are connected.
3)
If the green light on the Ethernet port does not light up, use the Optional Display to check that the unit’s Enable
DHCP is turned off and follow steps 4 through 8 If an optional display is not available, please call Hardy Tech
Support for further assistance.
4)
Using the unit’s Fixed IP address (the default is 192.168.000.100), the PC must now be assigned a unique IP
address (for example, 192.168.000.101)
There are two simple rules for the IP Address:
•
It must have the same network identifier as the computer.
•
It must have a different node identifier than the computer.
5)
On a PC running windows, open Internet Properties (TCP/IP).
Click in the ‘Use the Following IP Address’ checkbox; then enter the following into the TCP/IP Properties
dialog.
•
IP Address = 192.168.0.101
•
Subnet Mask = 255.255.255.000
6)
Select OK on the TCP/IP Properties dialog box. The computer is now configured.
7)
Enter the HI 6600 HGM’s IP address into the Windows PC Web browser to access the embedded web browser.
For example: http://192.168.000.100
NOTE
Additional information for setting up communications between the Gateway Module and the PLC,
DCS, or PC is under Network Communications (Chapter 6).
Summary of Contents for HI 6600-EIP
Page 16: ...HI 6600 Series Modular Sensor System User Guide Page 16...
Page 44: ...HI 6600 Series Modular Sensor System User Guide Page 44...
Page 97: ...HI 6600 Series Modular Sensor System User Guide Page 97 A ElectricalandMechanicalReview...
Page 107: ...HI 6600 Series Modular Sensor System User Guide Page 107 H Mechanical Installation...
Page 108: ...HI 6600 Series Modular Sensor System User Guide Page 108 J Electrical Inspection...
Page 109: ...HI 6600 Series Modular Sensor System User Guide Page 109 K InstallationCheckPoints...
Page 128: ...HI 6600 Series Modular Sensor System User Guide Page 128...