16. pH MEASUREMENT
The
HI84531
can be used as a pH meter for direct measurements.
Verify that the instrument has been calibrated before taking pH measurements. Set the instrument to
pH meter. From titrator mode press
Meter
until pH units are displayed.
If an electrode calibration hasn’t been performed, or the number of days exceeds the calibration time
out value set, the message “CAL DUE” will blink on the left side of the display (see Calibration Timeout
option in SETUP MENU section for details).
If “CAL DUE” is displayed, perform an electrode calibration.
Press
MENU
to access the instrument’s menu.
Press
HELP
to view the contextual help, every time you need additional information.
Press
STIR
to start/stop the stirrer.
Press
Titrator
to enter titration mode.
Press
CAL
to access the calibration menu.
Press
LOG
to save the current reading. A message indicating the free log space will be displayed for
a few seconds.
In order to take pH measurements, follow the next steps:
• Submerge the tip of pH electrode 2 cm (0.8’’) and the temperature probe into the sample to be
tested and stir gently. Allow time for the electrode to stabilize. When the reading becomes stable,
the (unstable measurement) symbol will disappear.
• If the pH reading is less than -2.00 pH or greater than 16.00 pH, the closest full-scale value will
be displayed blinking.