34
PU
m
P C
ALIB
rA
TION
Pr
OCEDU
rE
• At the end of the calibration, ”Calibration Completed“ appears on display. To repeat the calibration
press
Restart
or
ESC
to return to the main screen.
Notes:
If temperature probe is not connected, Manual Temperature Compensation is used and MTC
appears on the right side of the screen. If Automatic Temperature Compensation is in use the ATC
appears on the right side of the screen.
If an erroneous situation is encountered during the calibration, an error message is displayed and
the calibration can be restarted by pressing Restart. Prepare a new standard, rinse electrode,
temperature probe and dosing tip and try again.
If the calibration doesn’t complete and the max titrant volume of titrant is reached an error
message will be displayed. The calibration can be restarted by pressing Restart. Prepare a new
standard, rinse electrode, temperature probe and dosing tip and try again.
This error message appears when the pH reading exceeds the acceptable input limits (-2.00 <
pH > 16.00).
This screen appears when the stirrer is not working properly. Check the stir bar and beaker
content. Press Restart to try again.