33
PU
m
P C
ALIB
rA
TION
Pr
OCEDU
rE
• Press
Start
, wait for the syringe refill.
• Place the stir bar in the calibration solution beaker and put the beaker in the mini titrator top.
• Place the probe holder on the top of the beaker and secure it by turning clockwise.
• Rinse the pH electrode with deionized water and immerse into the calibration solution until the
reference junction is completely submerged. Be sure that the tip of the electrode is not hitting the stir
bar. If necessary, additional distilled or deionized water can be added.
• Verify if the titrant corresponds to selected titration range.
• Insert the dosing tip into the titrant tube sleeve. It is critical that the tip be immersed approximately
0.25 cm (0.1”) into the solution being titrated.
• Press
Continue
to begin the calibration and
Stop
to abort it.