38
TIT
rA
TION
Pr
OCEDU
rE
• The titration curve can be visualized during a titration by pressing
Plot ON
. Press
Plot OFF
or
ESC
to exit this mode.
• At the end of the titration, the total or strong alkalinity of water sample in the selected unit of
measurement (mg/L or meq/L) is displayed. The titration curve can be viewed by pressing
Plot ON
.
Press
Plot OFF
or
ESC
to exit this mode.
• Press
LOG
to record the concentration value into the instrument’s memory. A message will be
displayed for a few seconds indicating the amount of the free log space. Up to 200 log samples can
be recorded in the instrument’s memory.
• Press
Restart
to begin a new titration or
ESC
to return to the titration menu.
• If the concentration exceeds the range limits, the exceeded range limit will be displayed blinking.
Another titration can be started by pressing
Restart
.
• “Wrong input” error message appears when the input reading (pH, temperature) exceeds the
specified limits. The pH or temperature value and concentration will blink indicating an error.