15
SETUP
mENU
From your knowledge of expected concentrations, use the table below to determine which settings and
solutions are appropriate:
Measurement Range
Low
High
Range
mg/L
30.0 - 400.0
300 - 4000
meq/L
0.6 - 8.0
6.0 - 80.0
Acidity Type
Total
Strong
Total
Strong
pH Endpoint
4.5 pH
8.3 pH
4.5 pH
8.3 pH
Minimum sample pH
> 4.5 pH
> 8.3 pH
> 4.5 pH
>8.3 pH
Calibration Timeout
Disabled or 1 to 7 days
Set the number of days before the pH calibration expired warning message is displayed.
Press
Modify
to access the calibration timeout screen.
Use the
ARROW
keys to select the value.
Press
Select
to confirm or
ESC
to return to the setup menu without saving the changes.
pH Resolution
0.1 or 0.01
Press the displayed virtual option key to change the resolution.
Temperature Unit
°C, °F or K
Press the virtual option key to change the option.